ZNB-XAII Idapo fifa
FAQ
Q: Njẹ ẹrọ ṣiṣi silẹ fifa soke?
A: Bẹẹni, Universal IV ṣeto le ṣee lo pẹlu fifa idapo wa lẹhin isọdiwọn.
Q: Ṣe fifa soke ni ibamu pẹlu Micro IV Set (1 milimita = 60 silẹ)?
A: Bẹẹni, gbogbo awọn ifasoke wa ni ibamu pẹlu IV Set ti 15/20/60 dorps.
Ibeere: Ṣe o jẹ ẹrọ fifa peristaltic?
A: Bẹẹni, curvilinear peristaltic.
Q: Kini iyatọ laarin PURGE ati iṣẹ BOLUS?
A: Purge ti a lo lati yọ afẹfẹ kuro ni ila ṣaaju idapo.Bolus le ṣe abojuto fun itọju idapo lakoko idapo.Mejeeji purge ati oṣuwọn bolus jẹ siseto.
Awọn pato
Awoṣe | ZNB-XAII |
Fifa Mechanism | Curvilinear peristaltic |
IV Ṣeto | Ni ibamu pẹlu IV tosaaju ti eyikeyi bošewa |
Oṣuwọn sisan | 1-1500 milimita / h (ni awọn afikun 0.1 milimita / wakati) |
Purge, Bolus | 100-1500 milimita / h (ni awọn afikun 0.1 milimita / wakati) Yọọ nigbati fifa soke duro, bolus nigbati fifa ba bẹrẹ |
Yiye | ± 3% |
* Thermostat ti a ṣe sinu | 30-45 ℃, adijositabulu |
VTBI | 1-20000 milimita (ni awọn afikun 0.1 milimita) |
Ipo idapo | milimita / h, ju / min, akoko-orisun, iwuwo ara, ounjẹ |
Oṣuwọn KVO | 0.1-5 milimita / h (ni awọn afikun 0.1 milimita / wakati) |
Awọn itaniji | Idaduro, air-in-line, ṣiṣi ilẹkun, eto ipari, batiri kekere, batiri ipari, Agbara AC kuro, aiṣedeede mọto, aiṣedeede eto, imurasilẹ |
Afikun Awọn ẹya ara ẹrọ | Iwọn infused gidi-akoko, iyipada agbara adaṣe, bọtini odi, wẹ, bolus, iranti eto, akọọlẹ itan, titiipa bọtini, ile ikawe oogun, rotari koko, ayipada sisan oṣuwọn lai idekun fifa |
Oògùn Library | Wa |
Occlusion ifamọ | Ga, alabọde, kekere |
Itan Wọle | 50000 iṣẹlẹ |
Atẹgun-ni-ila erin | Oluwari Ultrasonic |
Alailowaya isakoso | iyan |
Ju Sensọ | iyan |
Agbara Ọkọ (Ambulansi) | 12± 1.2 V |
Ipese agbara, AC | 110/230 V (aṣayan), 50-60 Hz, 20 VA |
Batiri | 9.6 ± 1.6 V, gbigba agbara |
Igbesi aye batiri | 5 wakati ni 25 milimita / h |
Awọn iwọn otutu ṣiṣẹ | 10-30 ℃ |
Ọriniinitutu ibatan | 30-75% |
Afẹfẹ Ipa | 860-1060 hpa |
Iwọn | 130 * 145 * 228 mm |
Iwọn | 2.5 kg |
Ailewu Classification | Kilasi Ⅰ, tẹ CF |
![ZNB-XAII--fifun idapo (1)](http://www.kellymedgroup.com/uploads/ZNB-XAII-infusion-pump-1.jpg)
![ZNB-XAII--fifun idapo (2)](http://www.kellymedgroup.com/uploads/ZNB-XAII-infusion-pump-2.jpg)
![ZNB-XAII--fifun idapo (3)](http://www.kellymedgroup.com/uploads/ZNB-XAII-infusion-pump-3.jpg)
![ZNB-XAII--fifun idapo (4)](http://www.kellymedgroup.com/uploads/ZNB-XAII-infusion-pump-4.jpg)
![ZNB-XAII--fifun idapo (5)](http://www.kellymedgroup.com/uploads/ZNB-XAII-infusion-pump-5.jpg)
![xqt (8)](https://cdn.globalso.com/kellymedgroup/xqt-8.jpg)
![xqt (7)](https://cdn.globalso.com/kellymedgroup/xqt-7.jpg)
![xqt (5)](https://cdn.globalso.com/kellymedgroup/xqt-5.jpg)
![xqt (6)](https://cdn.globalso.com/kellymedgroup/xqt-6.jpg)
![xqt (4)](https://cdn.globalso.com/kellymedgroup/xqt-4.jpg)
![xqt (3)](https://cdn.globalso.com/kellymedgroup/xqt-3.jpg)
![xqt (1)](https://cdn.globalso.com/kellymedgroup/xqt-1.jpg)
![xqt (2)](https://cdn.globalso.com/kellymedgroup/xqt-2.jpg)
Kọ ifiranṣẹ rẹ nibi ki o si fi si wa